With an extended measurement range from micro-amps to hundreds of amps and milli-volts to kilovolts, the Powertek PA2801 and PA2802 are ideal power analyzers for appliance testing, standby power or Energy Star conformance testing.  High speed digitization of voltage and current allows accurate power analysis of both clean and distorted  power signals.  The graphical display includes numeric results, voltage and current waveform plots along with settings.  Measurements can be displayed or read via the communication ports; RS-232, GPIB, USB, and Ethernet, or sent directly at full resolution to a USB printer.

 Power Analyzer Features

  • Accurate with all AC/DC waveform types
  • High performance-to-cost ratio
  • Measures, W, VA, VAr, V, A, PF crest factors and harmonics
  • Base accuracy <0.08%. Current and voltage accuracies specified to less than 1mArms and 1Vrms respectively (<0.2%)
  • 2000Vpk and 150Apk measurable internally (external current probe capable)
  • Integrated line switch and inrush waveforms
  • Graphics display shows numerical results, waveforms, bar graphs, startup & history charts with zoom & scroll features
  • Wiring loss and voltage burden compensation
  • Harmonic Analysis
  • DC charge and discharge measurements
  • Frequency Range: DC and 20mHz – 200kHz
  • Communications ports include USB, GPIB, and RS-232 standard, plus optional Ethernet*
  • USB Flash drive support for data logging*
  • Provides PASS/FAIL tests to user limits*

 *Planned features coming soon



PA2801 PA2802 power analyzer rear panel input connection and interfaces


MEASUREMENT FUNCTION PA2801/PA2802
  Measurement Bandwidth Range DC, 0.02Hz to 200kHz
 VOLTS Vrms
 Refer performance graphs operation manual
Display Resolution 5 digit
Measurement Range 700mVrms to 850Vrms (2000Vpk)
(200mVrms with display blanking off)
Vpeak Ranges 850V (2000Vpk), 450V (675Vpk), 135V (200Vpk), 50V (80Vpk), 20V (30Vpk)
Max input volts 2000Vpk
Accuracy 0.08% Reading + 0.1% Range 0.04%/kHz
DC 0.2% Reading +0.2% Range + 10mV
Common Mode 90dB
Input Impedance 1Mohm 10pF
Current Arms
Refer performance graphs operation manual
Display Resolution 5 digit
Measurement Range 0.01Arms to 30Arms (150Apk)
Transducer (Ext Shunt) input 2mVpk to 15V (35Vpk).  Max input 70Vrms
Amps Ranges 65A (150Apk), 20A (50Apk), 6.5A (15Apk), 2A (5Apk), 0.65A (1.5Apk), 300mA (800mApk), 100mA (250mApk), 30mA (80mApk), 10mA (25mApk)
Max input amps 30Arms continuous 150Apk
Accuracy 0.09% Reading + 0.1% Range 0.05%/kHz
DC 0.2% Reading + 0.2% Range
Common Mode 90db
Input impedance 0.015 Ohm - 650mA-65A
0.3ohm - 10mA-300A
Power active power watts Accuracy 0.1% Reading + 0.1% Range
VA apparent power Accuracy 0.1% Reading + 0.1% Range
VAR reactive power Accuracy 0.1% Reading + 0.1% Range
General Specifications
Interfaces Standard USB, GPIB, and RS-232 standard
Optional Ethernet with RS232/Ethernet adapter
Environment Operating Temperature 0° to +55° C
Storage Temperature -20°C to +70°C
Humidity 10% to 90% RH non-condensing
Dielectric Strength Inputs to case 2kV AC 50/60Hz for 1 minute
AC line input to case 2kV AC 50/60Hz for 1 minute
Power AC line input voltage 90V to 264V
AC line input frequency 45Hz to 450Hz
AC line input fuse 20mm 1AT
Power Consumption 14W/28VA
Mechanical Weight 7.5lbs (3.4kg)
Dimensions (W x H x D) 13.8in x 4.7in x 9.5in
(35.05cm x 11.94cm x 24.13cm)


  


PA Power Switch