DP-50 High Voltage Differential Probe

The DP-50 differential voltage probes are an important accessory for any application involved in power electronics and high voltage systems.  These probes allow isolated and floating measurments with any oscilloscope or data recorder, allowing the oscilloscope to remain grounded.  Using a DP50  Differential probe, differential and floating (isolated) measurements can be made on a variety of ac power systems like 460V/230V/110V ac input power supplies, chopped and switching power supplies, thyristor controllers, variable speed motor drives (PWM signals), motor inrush current and measurement across current shunts.  The isolated input is made via two 4mm/banana safety sockets. The BNC output allows easy connection to an oscilloscope or DAQ.  The DP isolation probes are powered from a battery or power supply.

DP-50 Differential Probe Features:
  • High voltage differential probe
  • Max. Input Voltage:  +/-6.5kV DC or 6.5kVp-p
  • Bandwidth:  DC-50MHz
  • Input Impedance 54Mohm and 1pF
  • x100, x200, x500, x1000 Attenuator
  • Max to ground voltage: 6.5kV CATIII
  • Output Impedance 50ohm
  • Used for isolation of oscilloscope and data loggers
  • Suitable for power electronic systems and EV electric vehicle power systems
  • Allows isolated floating safe measurements
  • Prevents ground loop currents between test equipments
DP-50 Differential Probe Specifications
Differential Voltage DC and AC Peak:
6.5kVp-p
Bandwidth (50ohm load -3dB):
50MHz
Common Mode Voltage DC / AC peak:
3.25kVp-p
Common Mode Voltage RMS CATII / CATIII:
3.25kVp-p
To Ground Maximum Voltage:
3.25kVp-p CATIII
Attentuation (Switchable):
x100, x200, x500, x1000
Input R (Each Input):
27Mohm >+2%
Input C (Each Input):
<2pF
Maximum Operation Voltage (DC / AC peak):
<650V x100, 1300V x200, 3.25kV x500, 6.5kV x1000
Common Mode Rejection Ratio (CMRR):
60Hz: >10000:1
100Hz: >1000:1
1MHz: >300:1
Noise (Into 50 ohm Load):
<2.0mVrms
Input Impedance (Between Input):
54Mohm and 1.25pF
Accuracy (at 20-30 degrees Celsius, 70% RH after 20 minutes):
<+/-2%
Maximum output Voltage:
6.5Vp-p
Power Source:
External 9V DC Adaptor
Other:
Over Range Indication

DP50 Datasheet
CE marked