GP1700 Gain Phase, Frequency Response and LCRZ Analyzer

GP1700 Frequency Response Analyzer

GP1700 - 4 U high format.
Ideal for ATE applications

About the GP1700 Gain Phase Analyzer

The GP1700 PAV frequency response analyzer is a versatile and powerful test and measurement instrument. Using 10uV - 100Vpk (600Vpk with supplied probes) isolated inputs with a powerful DSP based Discrete Fourier Transform (DFT) measurement, accurate phase sensitive and frequency selective measurements are possible. In particular the GP1700's gain, phase, power and LCRZ measurements offer good accuracy and noise rejection, necessary in most switching power supply power conversion, ultrasonic and EL backlight applications. Its unique ability to measure impedance parameters during actual live operating conditions, up to 600Vrms, can reveal hidden circuit properties - not possible with conventional LCR meters. The main applications for the GP1700 include assessment of synchros, resolvers, LVDT, gain phase margins and stability in closed loop control systems, ultrasonic equipment, analysis of inductors, response of active/passive filters, piezo devices and amplifiers. Using its frequency sweep capability to create Bode Plots, impedance plots, Q factor, reactance and Resr plots. Particularly suited to situations where the measured signal is not referenced to ground or buried in noise and distortion.

Additional capability
  • Fast dual channel Dc-1MHz isolated voltmeters
  • Phase angle meter / Vector voltmeter / Phase Angle Voltmeter (PAV)
  • Graphical display for direct screen plotting
  • Inductive component testing, transformer analyzer / wound component tester, capacitance, resistance, reactance and impedance
  • Wideband wattmeter
  • THD and harmonics for voltage and current
  • Selective level meter (measures amplitude at a particular frequency)
  • Voltage & current waveform display
  • Fast measurement speeds exceeding 300 measurements/second
  • RS232, USB, GPIB (option) interfaces, comprehensive PC software, NI LabView drivers, MS Excel compatible files, MS Excel Macros available
Specifications for GP1700 Gain Phase Analyser measurements
Channels / display
2 isolated / 5 digits resolution (0.01° Phase Angle)
Measurement
Total Vrms, Fundamental, Inphase, Quadrature, Frequency, Phase angle plus Null Meter
Voltage Input Ranges
300V, 100V, 30V, 10V, 3V, 1V, 300mV, 100mV, 30mV,10 mV or Autoranging
Phase Input Ranges
0.00° to -360°, 0.00° to +360° or 0 to ±180°
Frequency Range
DC to 1MHz
Common Mode Rejection Ratio (CMRR)
5Hz to 999.99Hz:
126dB
1kHz to 5kHz:
110dB
5kHz to 32 kHz:
100dB
32kHz to 64kHz:
91dB
Harmonic Rejection
80dB (even and odd)
Max input
±100V peak direct 100V peak from earth (600Vpk using supplied DP9030 differential probe)
Input impedance
1 M // 30pF (excluding. Leads)
Coupling
AC or AC+DC
Time constant
0.2s, 1.5s or 12s
Output generator (sinewave)
1mV to 8Vrms (12Vpk) at 100mA
Accuracy (Signal and Reference)
Frequency Voltage Rdg + Rng Phase Gain Ratio
10mHz to 2kHz ± 0.03% ± 0.03% ±0.005° ±0.02dB
2kHz to 5kHz ± 0.05% ± 0.05% ±0.01° ±0.03dB
5kHz to 20kHz ± 0.10% ± 0.10% ±0.03° ±0.04dB
20kHz to 32kHz ± 0.10% ± 0.20% ±0.15° ±0.06dB
32kHz to 54kHz ± 0.10% ± 0.80% ±0.25° ±0.08dB
54kHz to 100kHz ± 0.10% ± 1.20% ±0.50° ±0.12dB
Useful Links
Example of GP1700 gain phase analyzer screen shots
gp1700 screen shot gp1700 screen shot gp1700 screen shot gp1700 screen shot gp1700 screen shot
Ordering
  • GP1700 Upright with 110V/230Vac universal power
  • GP1700R 19" 2U rack with 110V/230Vac universal power
All models include
  • USB-RS232 converter
  • RS232
  • RS232 cable
  • CD software utilities
  • 2 x ISO input cables probes
  • output cable
  • quick guide sheets
  • user manual
  • calibration certificate
  • certificate of conformance
Options
  • GPIB IEEE-488
  • Option 01A Active LCRZ head inc Kelvin connectors
  • Option 01P Passive LCRZ head inc Kelvin connectors
  • Option 03A injection transformer
  • Option 03A-LF injection transformer