The MC133c-3F model are compact, one-housing, three phase electric, ac power calibrators.
The MC133c-3F model are compact, one-housing, three phase electric, ac power calibrators. It can supply the device under test with one phase accurate AC voltage and current with calibrated phase shift or DC voltage and current. Output voltage can be set in five ranges up to 280V and output current in six ranges up to 30A.
Simulated electric power can be set independently in each phase. Frequency can be set in range from 15Hz to 1kHz with resolution 0.001Hz to 0.01Hz. AC Power Calibrator equipped with 5 1/2 digital DC meter with DC ranges 20mA or 10V for direct calibrating of various types of power transducers.
Using supplied cable adapter the current outputs can be joined together offering one phase current range to 90A AC/DC.
MC133c-3F offers only SINE wave output signals. Harmonic and inter-harmonic distortion is not available in this model.
General Technical Data | |
---|---|
Voltage range | DC: 1-280V AC: 1-600Vrms |
Current range | DC: 5m-30A AC: 5m-30Arms |
Resolution | 5 1/2 digit |
Frequency | Range : 15Hz to 1000Hz Accuracy : 0.005% Resolution : 1mHz for ≤500Hz, 10mHz for >500Hz |
Distortion | voltage output : <0.05% current output: <0.1% |
Internal ranges | Voltage: 10V, 30V, 70V, 140V, 280V, 600V Current: 300mA, 1A, 2A, 5A, 10A, 30A |
Phase shift range | 0.000° to 359.99° |
Power factor | Range: -1.000 to +1.000 Resolution: 0.001 Accuracy: dPF = (1-cos(φ+dφ)/cos(φ) |
Total electric power range | 0.005VA to 8400VA |
Total electric energy range | 0.03VAs to 2333VAhr |
Time interval range | 1s to 10000s |
Time interval accuracy | 0.01% + 0.1s |
DC/AC Voltage | ||||||
---|---|---|---|---|---|---|
Range | % of value + % of range | Max. current (mA) |
||||
DC | 40 - 70Hz | 15 - 40Hz 70 - 1kHz |
||||
1 - 10V | 0.015 + 0.01 | 0.012 + 0.01 | 0.016 + 0.01 | 100 | ||
10 - 30V | 0.015 + 0.01 | 0.012 + 0.01 | 0.016 + 0.01 | 200 | ||
30 - 70V | 0.015 + 0.01 | 0.012 + 0.01 | 0.016 + 0.01 | 300/200 | ||
70 - 140V | 0.015 + 0.01 | 0.012 + 0.01 | 0.016 + 0.01 | 300/200 | ||
140 - 280V | 0.015 + 0.01 | 0.012 + 0.01 | 0.016 + 0.01 | 200/150 | ||
280 - 600V | N/A | 0.016 + 0.01 | 0.024 + 0.01 | 60/50 |
DC/AC Current | ||||||
---|---|---|---|---|---|---|
Range | % of value + % of range | Voltage (Vpk) |
||||
DC | 40 - 70Hz | 15 - 40Hz 70 - 1kHz |
||||
5 - 300mA | 0.0175 + 0.01 | 0.0175 + 0.01 | 0.021 + 0.02 | 8/5 | ||
0.3 - 1A | 0.0175 + 0.01 | 0.0175 + 0.01 | 0.021 + 0.02 | 8/5 | ||
1 - 2A | 0.0175 + 0.01 | 0.0175 + 0.01 | 0.021 + 0.02 | 8/5 | ||
2 - 5A | 0.0175 + 0.01 | 0.0175 + 0.01 | 0.021 + 0.02 | 5 | ||
5 - 10A | 0.021 + 0.015 | 0.021 + 0.015 | 0.028 + 0.02 | 5 | ||
10 - 30A | 0.0245 + 0.015 | 0.0245 + 0.015 | 0.035 + 0.02 | 5 |
Phase and Power Factor | |||
---|---|---|---|
Current | Accuracy dφ (°)3 | ||
15 - 70Hz | 70 - 400Hz | 400 - 1kHz | |
5 - 8mA | 0.4 | 0.4 | 1.0 |
8 - 100mA | 0.05 | 0.1 | 0.4 |
0.1 - 10A | 0.01 | 0.1 | 0.4 |
10 - 30A | 0.05 | 0.1 | 0.4 |
DC/AC Power Energy | ||||
---|---|---|---|---|
Set Current | % of value(Power Accuracy in common application) | |||
EU grid (230V,50Hz) |
US grid (115V,60Hz) |
Aircraft onboard (115V,400Hz) |
Ship onboard (440V,60Hz) |
|
100mA | 0.054% | 0.054% | 0.086% | 0.057% |
1A | 0.038% | 0.038% | 0.051% | 0.042% |
10A | 0.045% | 0.045% | 0.057% | 0.048% |
30A | 0.047% | 0.047% | 0.063% | 0.050% |
Active power (W): dP = √(dU2 + dI2 + dPF2 + 0.012) [%]
Reactive power (VAr): dP = √(dU2 + dI2 + dPFvar2 + 0.012) [%]
Apparent power (VA): dP = √(dU2 + dI2 + 0.012) [%]
Accessories (included): | |
---|---|
Power cable | 1 pc |
Option 10 Test cable banana-banana 1m 1000V/32A black | 3 pcs |
Option 11 Test cable banana-banana 1m 1000V/32A red | 3 pcs |
Option 12 Test cable banana-banana 1m 1000V/32A blue | 3 pcs |
Option 13 Test cable banana-banana 1m 1000V/32A yellow | 3 pcs |
RS232 cable 1.5m | 1 pc |
Spare fuse | 1 pc |
Option 133-01 High current adapter | 1 pc |
Operation manual, CD | 1 pc |
Test report | 1 pc |
Options: | |
---|---|
Option 140-50 | Current coil x25, x50 |
Current coil 0950 | multiplication factors x10, x25, x50 1500A maximum |
Current coil 151-25 | multiplication factors x25 3000A maximum active cooling |
Option 10 | Test cable banana-banana 1000V/32A,1m,black |
Option 11 | Test cable banana-banana 1000V/32A,1m,red |
Option 12 | Test cable banana-banana 1000V/32A,1m,blue |
Option 13 | Test cable banana-banana 1000V/32A,1m,yellow |
GPIB cable 2m | Double shielded 2m IEEE488 cable |
POWER | Application SW for electric power transducers calibration |
We use cookies to enhance your browsing experience, serve personalised ads or content, and analyse our traffic. By clicking "Accept All", you consent to our use of cookies.